The HI801 iris spectrophotometer is designed to work with five different cuvettes:
- 22 mm round
- 16 mm round
- 13 mm vial
- 10 mm square
- 50 mm rectangular
The meter is supplied with three cuvette adapters:
- adapter for 16 mm cuvette
- adapter for 10 mm square cuvette
- adapter for 13 mm vial
Note: The 22 mm round and 50 mm rectangular cuvettes do not require adapters. The cuvettes can be
To prepare the meter for the use of adapters:
- Open the meter’s lid.
- Select the adapter according to the cuvette type required for the selected method.
- Orient the adapter so that indexing mark is aligned with the indexing mark located inside the meter.
- Using light pressure, push the adapter down until it reaches the bottom of the meter’s holder.
- The meter is ready for use. Always utilize the selected adapter for both “Zero” and “Read” measurements as specified in the method instructions.