Designed for use in clinical, biochemistry, biogenetic, petrochemistry and water/wastewater laboratories; environmental protection and other fields of quality control
- Large sample compartment
- Single beam optical system
- Nominal 5 nm spectral bandpass
The two-line LED digital display reads % transmittance, absorbance, direct concentration and wavelength. Only one function is required to enter factor for direct concentration readout. The built-in RS-232C serial interface enables the instrument to be connected to a printer or computer. Sample compartment accepts 5 mm to 100 mm pathlength cuvettes and various accessories. Both models include a four-position 10 mm square cuvette holder, one set of four matched optical glass 10 mm square cuvettes, an RS-232C interface, an operator’s manual and a dust cover.
The 2100 is a VIS model and the UV-2100 is a UV/VIS model.
NOTE: the UV-2100 is also available with an analog wavelength selector, call for details.
Thomas No.: | 8409B10 | 8406B12 |
Spectral Slitwidth: | 5 nm | 5 nm |
Optical System: | Single Beam, Grating System 1,200 lines/mm | Single Beam, Grating System 1200 lines/mm |
Wavelength Range: | 325-1,000 nm (VIS) | 200-1,000 nm (UV/VIS) |
Wavelength Accuracy: | ±2 nm | ±2 nm |
Repeatability: | ±1 nm | ±1 nm |
Readability | 1 nm | 1 nm |
Stray Light: | ±0.3% T at 340 and 400 nm | ±0.3% T at 220 and 400 nm |
Photometric Range: | 0-125% T,- 0.1-2.5 A, 0-1,999°C (0-1,999°F) | 0-125% T, 0.1-2.5 A, 0-1999 °C (0-1999°F) |
Photometric Accuracy: | ±0.004A@0.5A% T | ±0.004A @ 0.5A% T |
Light Source: | Tungsten Halogen Lamp | Tungsten Halogen/Deuterium Lamp |
Data Output: | Two-line, 9-LED, displays, Trans., Abs., Conc. and wavelength, RS-232C, Analog Output | Two line, 9-LED displays, Trans., Abs., Conc. and wavelength, RS-232C, Analog Output |
Dimensions (W x D x H): | 18.5″ x 15.7″ x 5.5″ (470 x 400 x 140 mm) | 18.5″ x 15.7″ x 5.5″ (470 x 400 x 140 mm) |
Weight: | 10 kg | 12 kg |
Power Requirements: | 115/230 Volt. ±10%, 60/50 Hz Adjustable | 115/230 Volt. ±10%, 60/50 Hz Adjustable |
Thomas No | Mfr No | Description |
---|---|---|
8406B24 | S-90-309Q | SQUARE CUVETTE, Quartz, 10mm pathlength, 3.5ml capacity, Pair |
8406B28 | S-90-311Q | CUVETTE, Long Path, 50 mm, Pair |