The General Purpose Penetration Pt100 Probe – HI768P is designed to measure the temperature of semi-solid and liquid samples.
The operating principle of resistance thermometers is based on the increase of electric resistance of metal conductors (RTD: Resistance Temperature Detectors) with temperature. Among the various metals to be used in the construction of RTDs, platinum (Pt), a noble metal, is one that can measure temperatures throughout a wide range with a stable, linear behavior. The most common RTD sensor using platinum is the Pt100. Hanna’s Pt100 thermometers and probes use multi-wire technology in the connection cable to obtain high levels of accuracy in industrial and laboratory applications.
With a strong, corrosion-resistant, AISI 316 stainless steel construction, the HI768P probe is able to measure samples up to 350°C (622°F).

| Specification Name | Detail |
|---|---|
| SKU | HI768P |
| Accuracy (@25?C/77?F) | ±0.25°C (±0.5°F) ±3% of reading |
| Recommended Operating Temperature | -30 to 350°C (-22 to 622°F) |
| Response Time (90% of final value) | 3 seconds |
| Body Length | L 100 mm x dia 3 mm (3.9″ x 0.12″) |
| Temperature Sensor | yes |
| Body Material | AISI 316 stainless steel |
| Cable | polyurethane (PUR)/straight; 1 m (3.3″) |
| Recommended Use | air, gases |
| Connection | Pt-100 |