pH/ORP/ISE, EC/TDS/Resistivity/Salinity/Seawater Specific Gravity, Turbidity, DO, Temperature and Atmospheric Pressure
Rugged, waterproof and easy to use, the HI 9829 is the ideal meters for field measurements of lakes, rivers and seas. Both meters display 1 to 12 parameters simultaneously from up to 15 user selectable parameters.
Combined with one of the HI 76×9829 series probes, the HI 9829 can measure water quality parameters such as pH, ORP, conductivity, turbidity, temperature, ions ions ammonium, nitrate, chloride (NH4+, NO3-—N or Cl-), dissolved oxygen concentration, resistivity, TDS, salinity, and seawater specific gravity. Atmospheric pressure is measured for DO compensation.
This meter features a graphic, backlit LCD that scales digits to fit up to 12 parameters and allows full configuration of each parameter measured along with an on-screen graphing capability.M
HI 9829 incorporates comprehensive GLP features and the transferring of data via USB connection. The HI 98290’s HELP key displays context sensitive help. The alpha-numeric keypad offers a user friendly way to complete the input fields.
- Field replaceable ISO 7027 compliant turbidity sensor
- Ammonium, chloride and nitrate ISE’s
- Logging from probe or meter
- Fully customizable instrument, probe, sensors and measurement specifications
- Display from 1 to 12 parameters with font dimension adjustment
- Field replaceable sensors
- pH/ORP or pH, four electrode EC and galvanic DO sensors
- Auto-recognition of all sensors
- Rugged probe with stainless steel tip has a diameter under 2” for wells and pipes
- Fast Tracker™—Tag I.D. System simplifies periodic monitoring
- Features a built-in barometer for DO compensation
- Quick or independent sensor calibration feature
- Measurement check eliminates erroneous readings
- Logged data can be displayed as graphs
- Graphic LCD with backlight
- USB for PC connectivity
- Good Laboratory Practice feature with last five calibrations recorded
- Meter accepts both alkaline and rechargeable batteries
- Waterproof protection for meter (IP67) and probes (IP 68)
Range | pH/mV of pH Input | 0.00 to 14.00 pH/±600.0 mV |
ORP mV | ±2000.0 mV | |
Ammonium-Nitrogen | 0.02 to 200 ppm (as N) | |
Chloride | 0.6 to 200 ppm | |
Nitrate-Nitrogen | 0.62 to 200 ppm (as N) | |
Conductivity | 0 to 200 mS/cm (absolute EC up to 400 mS/cm) | |
TDS | 0 to 400000 mg/L or ppm (the maximum value depends on the TDS factor) | |
Resistivity | 0 to 999999 Ω•cm; 0 to 1000.0 kΩ•cm; 0 to 1.0000 MΩ•cm | |
Salinity | 0.00 to 70.00 PSU | |
Seawater σ | 0 to 50.0 st, s0, s15 | |
Turbidity | 0.0 to 99.9 FNU; 100 to 1000 FNU | |
Dissolved Oxygen | 0.0 to 500.0%; 0.00 to 50.00 ppm | |
Atm. Pressure | 450 to 850 mm Hg; 17.72 to 33.46 in Hg; 600.0 to 1133.2 mbar; 8.702 to 16.436 psi; 0.5921 to 1.1184 atm; 60.00 to 113.32 kPa | |
Temperature | -5.00 to 55.00°C; 23.00 to 131.00°F; 268.15 to 328.15K | |
Resolution | pH/mV of pH Input | 0.01 pH / 0.1 mV |
ORP mV | 0.1 mV | |
Ammonium-Nitrogen | 0.01 ppm to 1 ppm; 0.1 ppm to 200 ppm | |
Chloride | 0.01 ppm to 1 ppm; 0.1 ppm to 200 ppm | |
Nitrate-Nitrogen | 0.01 ppm to 1 ppm; 0.1 ppm to 200 ppm | |
Conductivity | manual: 1 µS/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm; 1 mS/cm; automatic: 1 µS/cm from 0 to 9999 µS/cm; 0.01 mS/cm from 10.00 to 99.99 mS/cm; 0.1 mS/cm from 100.0 to 400.0 mS/cm; automatic mS/cm: 0.001 mS/cm from 0.000 to 9.999 mS/cm; 0.01 mS/cm from 10.00 to 99.99 mS/cm; 0.1 mS/cm from 100.0 to 400.0 mS/cm | |
TDS | manual: 1 mg/L (ppm); 0.001 g/L (ppt); 0.01g/L (ppt); 0.1 g/L (ppt); 1 g/L (ppt); autorange scales: 1 mg/L (ppm) from 0 to 9999 mg/L (ppm); 0.01 g/L (ppt) from 10.00 to 99.99 g/L (ppt); 0.1 g/L (ppt) from 100.0 to 400.0 g/L (ppt); autorange g/L (ppt) scales: 0.001 g/L (ppt) from 0.000 to 9.999 g/L (ppt); 0.01 g/L (ppt) from 10.00 to 99.99 g/L (ppt); 0.1 g/L (ppt) from 100.0 to 400.0 g/L (ppt) | |
Resistivity | dependent on resistivity reading | |
Salinity | 0.01 PSU | |
Seawater σ | 0.1 σt, σ0, σ15 | |
Turbidity | 0.1 FNU from 0.0 to 99.9 FNU; 1 FNU from 100 to 1000 FNU | |
Dissolved Oxygen | 0.1%; 0.01 ppm | |
Atm. Pressure | 0.1 mm Hg; 0.01 in Hg; 0.1 mbar; 0.001 psi; 0.0001 atm; 0.01 kPa | |
Temperature | 0.01°C; 0.01°F; 0.01K | |
Accuracy @ 20°C | pH/mV of pH Input | ±0.02 pH / ±0.5 mV |
ORP mV | ±1.0 mV | |
Ammonium-Nitrogen | ±5% of reading or 2 ppm, whichever is greater | |
Chloride | ±5% of reading or 2 ppm, whichever is greater | |
Nitrate-Nitrogen | ±5% of reading or 2 ppm, whichever is greater | |
Conductivity | ±1% of reading or ±1 µS/cm, whichever is greater | |
TDS | ±1% of reading or ±1 mg/L, whichever is greater | |
Salinity | ±2% of reading or ±0.01 PSU, whichever is greater | |
Seawater σ | ±1 σt, σ0, σ15 | |
Turbidity | ±0.3 FNU or ±2% of reading, whichever is greater | |
Dissolved Oxygen | 0.0 to 300.0%: ±1.5% of reading or ±1.0% whichever is greater; 300.0 to 500.0%: ±3% of reading; 0.00 to 30.00 ppm: ±1.5% of reading or 0.10 ppm, whichever is greater; 30.00 ppm to 50.00 ppm: ±3% of reading | |
Atm. Pressure | ±3 mm Hg within ±15°C from the temperature during calibration | |
Temperature | ±0.15°C; ±0.27°F; ±0.15K |
Temperature Compensation | automatic from -5 to 55°C (23 to 131°F) |
Logging Memory from Meter | 44,000 records |
Logging Interval | 1 second to 3 hours |
Computer Interface | USB (with HI 929829 software) |
FastTracker™ TAG ID | yes |
Waterproof Protection | IP67 |
Environment | 0 to 50°C (32 to 122°F); RH 100% |
Power Supply | 1.5V alkaline C cells (4) / 1.2V NiMH rechargeable C cells (4), USB, 12V power adapter |
Dimensions | 221 x 115 x 55 mm (8.7 x 4.5 x 2.2”) |
Weight | 750g (26.5 oz.) |
Thomas No | Mfr No | Description |
---|---|---|
1198Q34 | HI9829-00041 | HI9829 with pH/ORP, EC, DO, 4 Meter Probe, Carrying Case, 115V |