Description
The HI983307 is supplied with a direct two pin probe and 2 m (6.6’) cable with a ½” thread for flow-thru mounting. The probe has a temperature sensor to automatically compensate against temperature changes from 5 to 50°C (41 to 122°F) with a ? of 2% per degree.
In the measurement mode, a red LED will warn the user in the event the reading is outside of the alarm interval. A front trimmer allows manual one-point calibration. The electrical circuitry is tightly sealed inside the water-resistant enclosure. This EC monitor can be installed anywhere quickly and easily with the casings molded eye. The 12 VDC power supply allows continuous monitoring over extended periods of time.
| Specification Name | Detail |
|---|---|
| SKU | HI983307 |
| EC Range | 0.00 to 9.99 mS/cm |
| EC Resolution | 0.01 mS/cm |
| EC Accuracy | ±2% F. S. |
| EC Calibration | manual, one-point, through trimmer with HI70039 |
| EC/TDS Temperature Compensation | automatic, 5 to 50°C (41 to 122°F) with ?=2% |
| Setpoint | 0.70 to 3.50 mS/cm |
| Alarm | red LED blinks when measured value differs from set point more than ±0.20 mS/cm |
| EC/TDS Probe | HI7632/2 with 2 m (6.6’) cable and ½” thread for in-line installation |
| Power Supply | 12 VDC adapter (included) |
| Environment | 0 to 50°C (32 to 122°F); RH max 100% |
| Dimensions | 86 x 110 x 43 mm (3.4 x 4.3 x 1.7’’) |
| Weight | 215 g (7.6 oz.) |
| Ordering Information | Supplied with HI7632/2 EC/TDS probe, calibration screwdriver, 12 VDC adapter and instructions. |
| Warranty | 1 year |