Feature:
1.Built-in circulating 3D constant temperature technology
Temperature stability time is less than 30 mins, ensuring the accuracy and reliability of test data.
2.The world’s leading full-range series of ion chromatographic columns
High efficiency, large capacity of the columns for detecting ions of varied compositions.
3. Self-Regenerating Electrolytic Micro-membrane Suppressor
High pressure resistance, small dead volume, highly responsive to signals.
4. Auto-range Conductivity Detector
It can directly detect the signal from ppb to ppm without adjusting the range; Only one conductivity detector can detect anions and cations.
5. Observatory intelligent workstation
Integrated control; With intelligent start-up, shutdown and maintenance functions; Compatible with a variety of instruments; Customized images.
Technical Parameter:
|
Model |
BK-IC100 |
|
|
IC Pump
|
Type |
High-pressure and low-pulse two-piston tandem advection pump |
|
Maximum Pressure |
35 MPa (PEEK) |
|
|
Pressure Display Accuracy |
≤0.1MPa |
|
|
Flow Stability |
(0.2-0.5) mL/min≤3%; (0.5-1.0) mL/min≤2%; >1.0mL/min≤2%; |
|
|
Allowable Deviation of Flow |
(0.2-0.5) mL/min , ±5%; (0.5-1.0) mL/min, ±3%; >1.0mL/min, ±2%; |
|
|
Flow Range |
0.001~9.999mL/min |
|
|
Pressure Pulse |
≤0.5% |
|
|
Flow Precision |
≤0.1% |
|
|
Flow Accuracy |
≤0.1% |
|
|
Manual Sample Injector |
Maximum Pressure |
35MPa |
|
External Diameter of the Tube |
1/16″ |
|
|
Contact Material of the Rotor |
PEEK |
|
|
Contact Material of Medium |
PEEK/Ceramics |
|
|
Column Heater
|
Operating Temperature Range |
20~60℃ (68~140℉) |
|
Controlling Temperature Accuracy |
±0.01℃ |
|
|
Allowable Deviation of Column Heater’s Temperature |
±1℃ |
|
|
Temperature Stability |
≤0.05℃/h |
|
|
Conduction Detection System
|
Type |
Constant temperature auto-range conductivity detector |
|
Cell Volume
|
≤0.8μL |
|
|
Detection Range |
0~35000μS/cm |
|
|
Detection Resolution |
≤0.0020nS/cm |
|
|
Output Voltage |
–6000~+6000mv(adjustable) |
|
|
Electronic Noise |
0.02nS |
|
|
Baseline Noise |
≤0.001μS/cm |
|
|
Baseline Drift |
≤0.02μS |
|
|
Operating Temperature Range |
Room temperature +5℃~60℃ (41 ~140℉) |
|
|
Controlling Temperature Accuracy |
±0.01℃ |
|
|
Temperature Compensation |
1.7 %/℃ |
|
|
Maximum Pressure |
10.0MPa |
|
|
Linear Range |
≥103 |
|
|
Instrument Linearity |
≥0.999 |
|
|
Quantitative Repeatability |
≤1.0% |
|
|
Qualitative Repeatability |
≤0.5% |
|
|
Minimum Detectable Concentration |
Cl-≤0.0005ug/mL; Li+≤0.001ug/mL; BrO3-≤0.001ug/mL; |
|
|
Flow System
|
Plastic Flow Path |
Made of PEEK materials |
|
Six-way Valve |
PEEK material, pressure 5000psi; Independent automatic collecting and flow function. |
|
|
Suppressor |
Type |
Self-Regenerating electrolytic micro-membrane suppressor |
|
Maximum Pressure |
6.0MPa |
|
|
Dead Volume |
<50μL |
|
|
Power Supply |
220V 50/60Hz(Standard);110V 50/60Hz (Optional);150W |
|
|
External Size(L*W*H*) |
325*400*500mm |
|
|
Net Weight |
21kg |
|