The built-in security features on MS-TS analytical balances ensure results are always valid. Operators receive a warning if any problems are encountered or tolerances are not adhered to. Corrective actions can be taken immediately, avoiding costly reworking later.
LevelControl, MinWeigh and automatic internal adjustment with proFACT support compliance with regulations. The ISOLog function maintains a record of all weighing relevant changes to the balance’s settings. For additional security, settings can be passcode protected.
The 7” extra-large color TFT touchscreen display is operable through cotton, silicon and rubber gloves. An intuitive user interface and 18mm high digits bring comfort to your daily tasks.
Built-in function simplifies balance leveling
The built-in LevelControl function issues a warning when the balance is not level and provides onscreen guidance to help you level the balance correctly within seconds.
MinWeigh function ensures process tolerances
The weight value remains red until the net sample is above the pre-programmed minimum value. A user-defined minimum weight can also be programmed into the balance to provide an additional safety factor.
Passcodes protect balance settings
Create a passcode to restrict modification of balance settings to authorized personnel only – functions that affect metrological performance are protected against accidental alteration.
Easy-cleaning features save time and effort
Remove, clean and replace all of the QuickLock draft shield glass panels in just a few simple steps, without tools and without moving the balance. The housing has smooth surfaces and rounded edges for easy cleaning.
Mettler products are intended for sale/shipment within the US only. International customers should contact their local Mettler distributor to place orders for Mettler products.
Thomas Nos. | 1188D83 | 1188D84 | 1188D85 |
---|---|---|---|
Limit values | |||
Maximum capacity | 120 g | 220 g | 220 g |
Readability | 0.1 mg | 0.1 mg | 0.1 mg |
Tare range (from … to) | 0 … 120 g | 0 … 220 g | 0 … 320 g |
Repeatability (nominal) (sd) | 0.1 mg (100 g) | 0.1 mg (100 g) | 0.1 mg (100 g) |
Linearity deviation (test load) | 0.2 mg (20 g) | 0.2 mg (50 g) | 0.2 mg (50 g) |
Eccentricity deviation (test load)1 | 0.4 mg (50 g) | 0.4 mg (100 g) | 0.4 mg (100 g) |
Sensitivity offset (test weight) | 0.8 mg (100 g) | 1 mg (200 g) | 1.5 mg (300 g) |
Sensitivity temperature drift 2 | 1.5 ppm/°C | 1.5 ppm/°C | 1.5 ppm/°C |
Sensitivity stability 3 | 2.5 ppm/a | 2.5 ppm/a | 2.5 ppm/a |
Typical values | |||
Repeatability (at low load) | 0.08 mg | 0.08 mg | 0.08 mg |
Linearity deviation | 0.06 mg | 0.06 mg | 0.06 mg |
Eccentric deviation (test load) 1 | 0.1 mg (50 g) | 0.12 mg (100 g) | 0.12 mg (100 g) |
Sensitivity offset (test weight) | 0.2 mg (100 g) | 0.24 mg (200 g) | 0.6 mg (300 g) |
USP minimum sample weight (5% load, k=2, U=0.10%) | 160 mg | 160 mg | 160 mg |
Minimum sample weight (5% load, k=2, U=1%) | 16 mg | 16 mg | 16 mg |
Settling time | 2 s | 2 s | 3 s |
Weighing pan dimensions (mm) | ø 90 | ø 90 | ø 90 |
- according to OIML R76
- in the temperature range 10 to 30°C
- Stability of sensitivity with proFACT self-adjustment switched on; s: seconds; a: year (annum); sd: standard deviation
Thomas No | Mfr No | Description |
---|---|---|
1188D83 | 30133522 | MS104TS Analytical Balance, 120g x 0.1mg |
1188D84 | 30133523 | MS204TS Analytical Balance, 220g x 0.1mg |
1188D85 | 30133524 | MS304TS Analytical Balance, 320g x 0.1mg |
1145Z94 | 30133538 | MS104TS/A Legal for Trade Analytical Balance, 120g x 0.1mg |
1145Z95 | 30133539 | MS204TS/A Legal for Trade Analytical Balance, 220g x 0.1mg |
1145Z96 | 30133540 | MS304TS/A Legal for Trade Analytical Balance, 320g x 0.1mg |