Panel-mounted pH Analog Indicator with Self Diagnostic Test

SKU : 9792bdb3c53d

HI8510 is a panel-mounted pH indicator designed for simplicity of use in a wide range of industrial process applications. Designed for easy and fast installation, the HI 8510 is provided with membrane keypads on the front panel, large display, and auto-diagnostic functions to check pH electrode and instrument status.

  • Automatic temperature compensation ensures fast, accurately compensated readings even during sudden temperature fluctuations.
  • Displays values on the easy to read LCD.
  • A removable, transparent splash-proof cover protects the front panel.
Pemesanan dapat melalui :
SKU 9792bdb3c53d Kategori :
Share this :

Description

Easy Installation

HI 8510 is provided with membrane keypads on the front panel, large display, and auto-diagnostic functions to check pH electrode and instrument status.

Easy pH Electrode Connection

Allows for quick and easy connection to any pH meter or transmitter.

Self-testing Diagnostics

Built-in autodiagnostic functions to enable the user to check and troubleshoot any malfunctions. The functions are made via front panel keys to isolate the cause of malfunction whether it is due to pH electrode contamination, internal offset circuit or the amplifier circuit.

Protective Cover

A removable, transparent splash-proof cover protects the front panel.

Recorder Terminals

To enhance troubleshooting and the ability to record data while monitoring, simply attach a recording device to the instrument’s 4 to 20 mA output contacts, conveniently located on the front panel, to obtain a copy of the results on demand.

Automatic Temperature Compensation

A 2-wire Pt100 can be connected to provide automatic temperature compensation of measurements.

 

 

 

Specification NameDetail
skuHI8510
pH Range0.00 to 14.00 pH
pH Resolution0.01 pH
pH Accuracy (@20°C/68°F)±0.02 pH (0 to 100 °C)/ ±0.05 pH (-20 to 0 °C) (electrode); ±0.5% (transmitter)
Temperature CompensationFixed or automatic with Pt100 from ?20 to 100 °C (?4 to 212 °F)
CalibrationOffset: ±2 pH with OFFSET trimmer;Slope: 80 to 110% with SLOPE trimmer
Recorder Output0?20 mA or 4?20 mA (isolated)
Power Output±5 Vcc; 150 mA max load for amplified electrodes
BacklightContinuous ON
Power Supply115 or 230 Vac; 60/50 Hz
Environment$-10 to 50°C (14 to 122°F); RH max 95% non-condensing
Enclosure/ Case MaterialBlack anodized aluminium body; front and back with ABS; transparent splash-proof front cover
Dimensions141 x 69 mm (5.6 x 2.7”)
Weightkg (2.2 lb.)