pH Electrode for Continuous Monitoring with DIN Connector

SKU : eaaba78a59b9

The HI1611D is a glass body, single junction pH electrode with a DIN connector. This electrode has a single ceramic junction in the outer reference cell and the pH sensing portion is made with high temperature glass.  This pre-amplified electrode features a built-in temperature sensor for temperature compensated measurements in a single probe design. This design consideration is ideal for measurements that require continuous monitoring. The HI1611D is recommended for use with Hanna’s HI8314 pH meter.

  • Built-In Temperature Sensor
  • High Temperature (HT) Glass
  • Gel Electrolyte
Pemesanan dapat melalui :
SKU eaaba78a59b9 Kategori :
Share this :

Description

 

Hanna Instruments offers a wide variety of pH electrodes that are designed for many different applications. The type of glass used for sensing pH, bulb shape, body material, type of junction, type of reference and electrolyte used are just some of the design considerations.

The HI1611D uses high temperature (HT) glass, spherical bulb, glass body, single ceramic junction, and gel electrolyte.

High-Temperature Glass Formulation

The measurement of pH at very high temperatures is detrimental to the sensitive glass bulb and will shorten the life of it. A pH electrode with general-purpose (GP) glass will have a resistance of 100 megaohms at 25°C while the resistance of HT glass is around 400 megaohms at 25°C. As the HI1611D is used at elevated temperatures the resistance decreases to approach the that of GP glass. The HI1611D is suitable to use with samples that measure from 0 to 80°C.

Spherical Bulb

The spherical bulb is for general purpose use. Other tip shapes include conic for penetration and flat tip for surface measurements.

Glass Body

The glass body is ideal for laboratory use. The glass is resistant to many harsh chemicals and is easily cleaned. The glass body also allows for a fast transfer of heat to the internal reference electrolyte. The mV generated by the reference cell is temperature-dependent. The faster the electrode reaches equilibrium, the steadier the reference potential.

Single Ceramic Junction

The outer junction of an electrode, also known as a salt bridge, is a necessary component of the electrical circuit. The movement of ions must flow through the junction for a steady reading. The outer reference has a single ceramic frit. The ceramic is a porous material that is easily fused with the glass body and has a similar expansion coefficient. A single ceramic junction has a flow rate of 15-20 ?L/hour. Other junction types are available with higher flow rates and made with different materials.

DIN Connector

The HI1611D uses a DIN connector. This type of connector is generally proprietary to the meters they are supplied with and may not be interchangeable. The HI1611D is recommended for use with Hanna’s HI8314 pH meter. Other types of connectors include BNC, screw type, T-type, and 3.5mm to name a few.

 

 

Specification NameDetail
SKUHI1611D
DescriptionpH electrode
Referencesingle, Ag/AgCl
Junction / Flow Rateceramic, single
Electrolytegel
Max Pressure2 bar
RangepH: 0 to 14
Recommended Operating Temperature0 to 80°C (32 to 176°F) – HT
Tip / Shapespheric (dia: 9.5 mm)
Temperature Sensoryes
Amplifieryes
Body Materialglass
Cable5-pole; 1 m (3.3’)
Recommended Usecontinuous monitoring
ConnectionDIN*
Special Note* Recommended for use with HI8314 pH meter