Refillable Double Junction pH Electrode with BNC Connector For Fluoride Applications

SKU : d8eda98f8c9b

The HI1143B is a glass body, refillable, double junction pH electrode with a BNC connector. This electrode has a single ceramic junction in the outer reference cell and the pH sensing portion is made with fluoride resistant glass. This design consideration is ideal for samples with fluoride.

  • Refillable Electrode
  • Double Junction Design
  • Fluoride Resistant Glass

Info penawaran harga dan pemesanan hubungi:

SKU d8eda98f8c9b Kategori :

 

Hanna Instruments offers a wide variety of pH electrodes that are designed for many different applications. The type of glass used for sensing pH, bulb shape, body material, type of junction, type of reference and electrolyte used are just some of the design considerations.

The HI1143B uses Fluoride Resistant (HF) glass, spherical bulb, glass body, single ceramic frit, double junction and is refillable with 3.5M KCl.

Fluoride Resistant Glass Formulation

The measurement of pH in samples containing fluroide is detrimental to the sensitive glass bulb and will shorten the life of it. HF resistant glass is designed for use in aggressive applications that have fluoride ions for extended life.

Spherical Bulb

The spherical bulb is for general purpose use. Other tip shapes include conic for penetration and flat tip for surface measurements.

Glass Body

The glass body is ideal for laboratory use. The glass is resistant to many harsh chemicals and is easily cleaned. The glass body also allows for a fast transfer of heat to the internal reference electrolyte. The mV generated by the reference cell is temperature-dependent. The faster the electrode reaches equilibrium, the steadier the reference potential.

Single Ceramic Double Junction Reference

The HI1143B is a double junction design. pH electrodes are available as single junction or double junction. See below for a complete description of the differences. The junction also known as a salt bridge is a necessary component of the electrical circuit. The movement of ions must flow through the junction for a steady reading. The outer reference has a single ceramic frit. The ceramic is a porous material that is easily fused with the glass body and has a similar expansion coefficient. A single ceramic junction has a flow rate of 15-20 ?L/hour. Other junction types are available with higher flow rates and made with different materials.

Refillable

The HI1143B is a refillable probe. Since it is a double junction pH electrode the fill solution is the HI7082 3.5M KCl. This solution does not contain any silver as with single junction electrode. The absence of silver will prevent any silver precipitate from forming at the junction surface and clogging it. Clogging of the junction will result in drifty and erratic readings.

BNC Connector

The HI1143B has a BNC connector. This type of connector is universal in that it can be used on any pH meter that has the female BNC probe input. Other types of connectors include DIN, screw type, T-type, and 3.5mm to name a few. These other types of connectors tend to be proprietary for a particular type of meter and are not interchangeable.

Single Junction Versus Double Junction pH Electrodes

electrode_fill_solutions_comparison

Conventional electrodes are normally single junction. As depicted by the figure above, these electrodes have only a single junction between the internal reference wire and the external solution. Under adverse conditions, such as high pressure, high temperature, highly acidic or alkaline solutions, the positive flow of the electrolyte through the junction is often reversed resulting in the ingress of sample solution into the reference compartment. If this is left unchecked, the reference electrode can become contaminated, leading to complete electrode failure. Another potential problem with single junction electrodes is the clogging of the junction due to silver chloride (AgCl) precipitation. Silver can be easily precipitate in samples that contain Tris buffer or heavy metals. When the electrolyte solution makes contact with the sample, some AgCl will precipitate on the external face of the junction. The result is drifty readings obtained from the sensor.

Hanna’s double junction system, as the name implies, has two junctions, only one of which is in contact with the sample as shown in the figure. Under adverse conditions, the same tendency of sample ingress is evident. However, as the reference electrode system is separated physically from the intermediate electrolyte area, the contamination of the electrode is minimized. The likelihood of clogging of the junction is also reduced with a double junction electrode since the outer reference cell uses a fill solution that is “silver-free.” Since there is no silver present, there is no precipitate that can form to clog the junction.

 

 

Specification NameDetail
SKUHI1143B
Descriptionrefillable, combination pHelectrode for fluoride applications
Referencedouble, Ag/AgCl
Junction / Flow Rateceramic, single / 15-20 µL/h
ElectrolyteKCl 3.5M
Max Pressure0.1 bar
RangepH: 0 to 10
Recommended Operating Temperature-5 to 60°C(23 to 140°F) — HF
Tip / Shapespheric (dia: 9.5 mm)
Temperature Sensorno
Amplifierno
Body Materialglass
Cablecoaxial; 1 m (3.3′)
Recommended Usesamples with fluoride (max 2 g/L @pH 2 and temperature < 60°C)
ConnectionBNC