pH Controller and Pump

SKU : 711e415b14b7

The BL7916 combines a pH meter with a chemical dosing pump in order to maintain a desired set point in process applications.

  • Two advanced instruments in one compact unit.
  • Convenient installation with all controls on front panel.
  • Widely used in plating, wastewater treatment, water treatment and swimming pools.

Info penawaran harga dan pemesanan hubungi:

SKU 711e415b14b7 Kategori :

Easy Installation

Designed with mounting holes built into a rugged base, BL7916 is simple to install. All of the controls and pump assemblies are conveniently located on the front of the unit.

Proportional Dosing

A proportional control slows down the stroke rate as the measured value approaches the user-selectable set points, avoiding overdosage of chemicals. This feature makes the pump’s dosing more accurate, saves chemicals and eliminates unnecessary and costly corrections to your process, especially with slow reacting chemicals.

Simple Pump Action

A positive displacement solenoid with few moving parts makes the BL7916 more reliable than motor-driven pumps since there is no rotating parts, gears or cams; drastically reducing any chance of mechanical failure.

Isolated Recorder Output

To enhance troubleshooting and the ability to record data while monitoring, simply attach a recording device to the instrument’s 4 to 20 mA output contacts, conveniently located on the front panel, to obtain a copy of the results on demand.

Simple Pump Action

A positive displacement solenoid with few moving parts makes the BL7916 more reliable than motor driven pumps since there is no rotating parts, gears or cams; drastically reducing any chance of mechanical failure.

Auxiliary Dosing Contacts

Allows the user to attach a mixer or priming pump that is activated only when the pump is dosing.

 

 

 

Specification NameDetail
skuBL7916-1
pH Range0.00 to 14.00 pH
pH Resolution0.01 pH
pH Accuracy (@20°C/68°F)±0.01 pH
Typical EMC Deviation±0.1 pH
Input Impedance10¹² Ohm
DosageProportional, acid or base, user selectable
Dosing ContactIsolated, 2A, Max. 240V, resistive load, 1,000,000 strokes
Alarm ContactIsolated, 2A, Max. 240V, resistive load, 1,000,000 strokes
CalibrationOffset: ±1 pH with trimmer
Slope: 85 to 115% with trimmer
Recorder Output4 to 20 mA (isolated)
Power Supply10/115V ±15%
50/60Hz (40W)
Environment0 to 50°C (32 to 122°F); RH max 95% non-condensing
Dimensions221 x 142 x 181 mm (8.7 x 5.6 x 7.1”)
WeightApproximately 5 kg (11 lb.)
Ordering InformationEach pump is supplied complete with: discharge and suction valves, LDPE tubing, 7 m (23′), power cord, and instruction manual.